8.
JEROME
631-X TECHNICAL SPECIFICATIONS
Range
Sensitivity
Precision
Accuracy
Response time-sample mode
10 to 50 ppm (Range 3)
1.0 to 10.0 ppm (Range 2)
0.10 to 1.00 ppm (Range 1)
0.001 to 0.100 ppm (Range 0)
Response time-survey mode
10 to 50 ppm (Range 3)
1.0 to 9.9 ppm (Range 2)
0.10 to 0.99 ppm (Range 1)
0.001 to 0.099 ppm (Range 0)
Flow rate
Power requirements
Fuse
Internal battery pack
- Standard model
- XE / 220V model
Operating environment
Case construction
Dimensions – standard model
Dimensions – XE model
Weight – standard model
Weight – XE model
Digital meter display
Certification
Customer Service 800-528-7411 or 602-470-1414
0.003ppm (3ppb) to 50ppm H
0.003ppm H
S
2
5% relative standard deviation
Range 0: ± 0.003ppm at 0.050ppm H
Range 1: ± 0.03ppm at 0.50ppm H
Range 2: ± 0.3ppm at 5.0ppm H
Range 3: ± 2ppm at 25ppm H
13 seconds
16 seconds
25 seconds
30 seconds
3 seconds
6 seconds
15 seconds
20 seconds
150 ± 10ml/min (0.15 ± .01 liters/min)
100-120 V~, 50/60 Hz, 1 A or 220-240 V~, 50/60 Hz, 1 A
F1A 250V, 5mm X 20mm
Rechargeable Nickel Cadmium
AMETEK BROOKFIELD P/N Z4000 0907
AMETEK BROOKFIELD P/N Z4000 0908
0° to 40 °C, non-condensing, non-explosive
Aluminum alloy
33 cm L x 15 cm W x 10 cm H (13" L x 6" W x 4" H)
35 cm L x 18 cm W x 18 cm H (14" L x 7" W x 7" H)
3.18 kilos (7 pounds)
3.5 kilos (8 pounds)
Liquid crystal display (LCD)
CE mark on 220-240 V~, 631-XE model only.
S in four graduated ranges
2
S
2
S
2
S
2
S
2