Instructions Regarding Use Of Remote Control - Otto Bock C-Leg 3C98-3 Instructions For Use Manual

Hide thumbs Also See for C-Leg 3C98-3:
Table of Contents

Advertisement

Available languages

Available languages

CAUTION
Safety mode cannot be deactivated
Falling due to unexpected behaviour of the product because of changed damping behaviour.
If you cannot deactivate safety mode by recharging the battery, a malfunction has occurred.
Do not continue using the defective product.
The product must be inspected by an authorised Ottobock Service Centre. The prosthetist
is your contact person.
CAUTION
Safety signal occurs (ongoing vibration)
Falling due to unexpected behaviour of the product because of changed damping behaviour.
Observe the warning/error signals (see Page 89).
Do not continue using the product after the safety signal has been emitted.
The product must be inspected by an authorised Ottobock Service Centre. The prosthetist
is your contact person.

4.9 Instructions regarding use of remote control

CAUTION
Improper use of remote control
Falling due to altered damping behaviour as a result of unexpected switching into MyMode.
Receive instruction in the proper use of the remote control unit.
CAUTION
Independent user changes or modifications made to remote control unit
Falling due to changed damping behaviour caused by unexpected switching to a MyMode.
Have any changes or modifications to the remote control carried out only by Ottobock
authorised, qualified personnel.
CAUTION
Incorrect mode switching using remote control
Falling due to unexpected behaviour of the product because of changed damping behaviour.
Ensure that you stand securely during all switching processes.
Verify the changed damping characteristics after switching, and observe the feedback from
the acoustic signal emitter and the display of the remote control.
Switch back to basic mode once the activities in MyMode have been completed.
NOTICE
Remote control damage due to water penetration (warranty lapses)
Malfunction of the remote control.
If necessary, let the remote control unit dry at room temperature (at least 1 day).
The remote control unit should be inspected by the prosthetist before it is used again. If
necessary, the prosthetist will send the unit to an authorised Ottobock Service Centre.