Error after disconnecting the battery charger (error on self-test)
Beep
Vibration
signal
signal
3 x short
3 x short

14.2.3 Error messages while establishing a connection

Error message
DEVICE NOT FOUND
DEVICE DISCONNEC­
TED
Error
Minor error:
e.g. service interval has been
exceeded, disruption of a sensor
signal
Cause
A previously saved con­
nection to the prosthesis
could not be estab­
lished.
A current connection to
the prosthesis was inter­
rupted
Resolution
Contact Ottobock
Correction
Check the following points:
Distance from the prosthesis to the
remote control
Battery charge level of the prosthesis
Active connection with the adjustment
software?
Bluetooth of the prosthesis switched
on? (see Page 83)
Prosthesis switched on? (Switching off
the product)
If multiple prostheses were stored in
the remote control, was the correct
prosthesis selected?
1) Connect the battery charger to the
prosthesis, then charge the battery for
at least 3 hours. You can try to estab­
lish connection again while the battery
is charging.
2) Use the  key to confirm the error mes­
sage, then try to establish the connec­
tion again.
3) You will find more information in the
section "Managing prostheses" (see
Page 68).
Check the following points:
Distance from the prosthesis to the
remote control
Was the prosthesis connected to anoth­
er remote control?
The connection that was already saved
in the remote control must be deleted in
order to re-connect the prosthesis to
this remote control. A new connection
from the prosthesis to this remote con­
trol must then be established (see
Page 68).
Use the  key to confirm the error mes­
sage, then try to establish the connec­
tion again.