Error Messages While Establishing A Connection With The Cockpit App; Status Signals - Otto Bock 1B1-2 Meridium Instructions For Use Manual

Hide thumbs Also See for 1B1-2 Meridium:
Table of Contents

Advertisement

Appendices
LED on
LED on
power sup­
battery
ply
charger
Beep signal
4 x short at intervals of
approx. 20 sec. (continu­
ously)
Error after disconnecting the battery charger (error on self-test)

15.2.3 Error messages while establishing a connection with the cockpit app

Error message
Component was connec­
ted
to
another
device.
Establish connection?
Mode change failed

15.2.4 Status signals

Battery charger is connected
LED on
LED on
power sup­
battery
ply
charger
Battery charger disconnected
Beep sig­
Vibration
nal
signal
1 x short
1 x short
3 x short
36
Error
Battery is fully charged (or connection
with product is interrupted).
Error
Charging the battery outside the allowable
temperature range
Cause
The component was connec­
ted to another device
An attempt was made to
switch to a different MyMode
while the component was in
motion (e. g. while walking)
A current connection to the
prosthesis was interrupted
Event
Power supply and battery charger operational
Result
Self-test completed successfully. Product is operational.
Maintenance note
Conduct the self-test again by connecting/disconnecting the battery charger. If the
beep signal sounds again, product maintenance should be carried out by an author­
ised Ottobock Service Center.
The product can be used without restrictions. However, vibration signals may not be
generated.
Resolution
heard.
If no signal is emitted, the connection to
the product is interrupted.
If the connection to the product is inter­
rupted, an authorised Ottobock Service
Centre must inspect the product, battery
charger and power supply.
Resolution
Check whether the specified ambient con­
ditions for charging the battery are met
(see Page 32).
Correction
To disconnect the original connection, tap the "OK"
button.
If the original connection is not to be disconnected,
tap the "Cancel" button.
For safety reasons, switching MyModes is only per­
mitted when components are at rest, e.g. while
standing or sitting.
Check the following points:
Distance from the prosthesis to the device
Charge level of the prosthesis
Bluetooth on the prosthesis switched on? (see
Page 28)
Hold the component with the sole of the foot
facing up to make the component "visible" for
2 minutes.
Prosthesis switched on? (see Page 28)
If multiple prostheses were stored, was the cor­
rect prosthesis selected?
1B1-2=* Meridium