3) By pressing the key, Bluetooth is either turned on permanently (ON) of turned off (OFF).
If Bluetooth is to be switched off, the remote control must be switched off after the key has
been pressed.
4) To leave the menu without saving changes, use the
confirm this with the key.
9.4.2 Switching Bluetooth off/on using remote control app
Switching off Bluetooth
1) Once the app has been started, tap the
Bluetooth of the component.
2) Follow the on-screen instructions.
Switching on Bluetooth
1) Rotate the component or connect/disconnect the battery charger.
→ Bluetooth is switched on for approx. 2 minutes. During this time, the app must be started
in order to establish a connection to the component.
2) Follow the on-screen instructions.
→ If Bluetooth is switched on, the
10 Additional operating states (modes)
10.1 Empty battery mode
Beeps and vibration signals are emitted if the available battery charge level is less than 5% (see
Page 89). During this time, damping settings are set to their safety mode values. The prosthesis is
then switched off. You can switch back to basic mode (mode 1) from empty battery mode by
charging the product.
10.2 Mode for charging the prosthesis
The product is non-functional during charging.
The product is set to safety mode damping. This may be low or high depending on the setting in
the adjustment software.
10.3 Safety mode
The product automatically switches to safety mode if a critical system error occurs (e.g. failure of
a sensor signal) or when the battery is empty. Safety mode remains in effect until the error has
been rectified.
Default damping values are activated in safety mode. This makes limited walking possible for the
user even though the system is not active.
The switch to safety mode is indicated by beeps and vibration signals immediately prior to switch
ing (see Page 89).
Safety mode can be disabled by connecting then disconnecting the battery charger. If the
product switches into safety mode again, this means a permanent error exists. The product must
be inspected by an authorised Ottobock Service Centre.
10.4 Overheating mode
When the hydraulic unit overheats due to uninterrupted, increased activity (e.g. extended walking
downhill), damping is increased along with the rising temperature in order to counteract the over
heating. When the hydraulic unit cools down, the product switches back to the damping settings
that existed before the overheating mode.
Overheating mode is indicated by a brief vibration every 5 seconds.
84 | Ottobock
symbol in the main menu to switch off the
symbol appears on the screen.
key to select the
symbol and