→ A scale will appear in the display of the remote control and shows the value currently set
for the parameter as well as the parameter's boundary values (see fig. 35).
3) Use the ,
keys to change the parameter setting and confirm this with the key.
To leave the menu without saving changes, use the ,
confirm this with the
→ The message "SUCCESSFUL" appears briefly to confirm these changes were saved in
the prosthesis.
4) If necessary, switch off the remote control by selecting and confirming the menu option "RC
POWER OFF".
If you need to change the settings of another mode, you must first switch into that mode.
9.3.2 Using remote control app to change prosthesis setting
1) Once the component is connected and in the desired MyMode, tap the
menu.
→ A list appears with the parameters for the currently active mode.
2) Change the setting of the desired parameter either by tapping the "<", ">" symbols or moving
the slider control.
If you need to change the settings of another mode, you must first switch into that mode.
9.3.3 Overview of adjustment parameters in basic mode
The parameters in basic mode describe the dynamic behaviour of the prosthesis in a normal gait
cycle. These parameters act as basic settings for automatically adjusting the damping behaviour
to the current motion situation (e.g. ramps, slow walking speed, etc.).
The stance function and/or the sitting function can also be activated/deactivated. Further informa
tion on the stance function (see Page 76). Further information on the sitting function (see
Page 77).
82 | Ottobock
key.
keys to select the symbol
symbol in the main
and