Setting The Device Properties For Digital Input Modules - Ametek FlarePro User Manual

Process mass spectrometer
Table of Contents

Advertisement

Setting the Device Properties for Digital Input Modules

1.
From the 2000 Software, click the Edit menu and select Device
Properties (Figure 6-2). Select your Opto 22 SNAP I/O System device
from Select a device to edit. Click OK. The Device Properties dialog
box (Figure 6-3) appears.
If a digital device is not listed, click Cancel. On the main screen click
the Edit menu and Add Device. Click Optomux Digital. Once the
digital module has been added, you can proceed with setting the
parameters.
Figure 6-2.
Device Properties
screen.
2.
From the General tab configure the Communications parameters as
follows:
• Bits per Second:
• Address:
• Type:
Figure 6-3.
General tab, Device
Properties screen.
3.
Make sure Device enabled and Live data boxes are checked.
57,600
128
Optomux Digital
Older I/O systems had a single digital input or output per module.
The SNAP I/O system combines four inputs or outputs into one
module. For the SNAP I/O system, the Process 2000 module
number refers to a particular digital I/O channel, not the physical
module. For example, if a single digital output module is installed
in the first SNAP rack position, the Process 2000 will refer to it as
Modules 0, 1, 2 and 3. See B3000 I/O Mapping and Example in
this chapter.
Opto Snap I/O Data Acquisition System | 6-11