Replacing the Scan Board

The Scan board is located across from the Control board and to its right,
behind the LED panel and the front heat sink. It is the longer of the two
boards attached to the LED panel. See Figure 5-1.
To remove the Scan board:
1.
Remove the I/O board.
2.
Remove the 1/4" screws from the front top and bottom casing of the
unit and the 3/16" screw from the right side (Figure 5-1).
3.
Loosen or remove the screws at the top and bottom of the option plate
to allow easier access to the removal of the Control board.
4.
Grab the heat sink and pull on it to remove the assembly with the
Control and Scan boards. It is connected to the unit by four connectors.
5.
Once the assembly has been removed from the unit casing, remove
the upper left and bottom right 3/8" screws with internal tooth lock
washer from the front heat sink.
6.
Gently pull on the heat sink to remove it. The LED panel is now
visible.
7.
Remove the 1/4" screw from the top right of the LED panel.
8.
Gently pull the Scan board away from the adhesive gap pad on
the back of the LED panel.
To replace the Scan board:
1.
Attach the LED panel at the upper right using the 1/4" screw.
2.
Replace the front heat sink using the 3/8" screws and lock washers on
upper left and bottom right.
3.
Insert the board into the unit making sure the four connectors (2 on
each board) are engaged.
4.
Replace or retighten the screws for the option plate to put it back in place.
5.
Replace the 1/4" screws on the top and bottom and the 3/16" screw on
the right side of the unit casing.
6.
Replace the I/O board.
There are adhesive gap pads on the back side of the front heat sink
and the LED panel to ensure that all fixtures are secure and in
place.
Board Replacement | 5-7