Disassembling the Enclosed Source
Figure 4-2.
Closed/enclosed
analyzer head.
|
4-14
FlarePro Process Mass Spectrometer
Remove the Filament (see instructions in this chapter).
Loosen the set screws on the Barrel Connectors that secure the
three leads coming down from the Source Elements. The
filament bases can be left intact. Remove the Source Assembly
from the Mass Filter.
1. Loosen and remove the three (3) locknuts that hold the source
connection rods in place on the lenses. Refer to Figure 4-2.
2. Loosen and remove the three (3) screws that hold the lens
together in sandwich form. Be careful not to lose the small
ceramic or metal washers from the source; each serves to hold the
source in precise alignment.