Figure 3-2.
Power Input Pinouts.
4.
Attach the communications cable to the RS-232 port on the back of the
electronics.
5.
Set the RS-232 / RS-485 switch on the back of the electronics to RS-232.
6.
Make sure that the power cable connection on the back of the
electronics is secure and plugged into the power panel.
7.
Connect the cable for the status LEDs.

Connecting the Electronics to the DC Line Power

Before connecting the 24 VDC power to the electronics, ensure
that the electronics are connected to the sensor. Otherwise, damage
to the electronics is possible.
The electronics unit requires 24 volts DC (± 1 volt) and 3.3 amps.
1.
Connect the DIN connector from the DC power panel inside the
enclosure to the power-input connection on the back of the electronics
unit.
2.
Align the key and push in the connector.
3.
Push the outer ring over the connector and rotate clockwise until the
ring locks in place (Figure 3-2).
PIN 1 = +24VDC
PIN 2 = +24VDC
PIN 3 = 24V RETURN
PIN 4 = 24V RETURN
POWER
INPUT
KEY