Electronics Installation and In-line Connections

Figure 3-1.
Electronics Pinouts.
3-2 | FlarePro Process Mass Spectrometer
The electronics are shipped in a separate package along with the
FlarePro. The electronics will be factory-configured with the I/O
board installed and the appropriate cables ready to be installed in
the FlarePro enclosure.

Connecting the Electronics to Quadrupole Head

Power must be off.
Always remove 24 VDC power from the AMETEK electronics
before removing the electronics from the sensor.
1.
Align the key on the electronics unit to the notch on the collar end of
the quadrupole head assembly.
2.
Insert the electronics onto the quadrupole head until firmly seated (do
not twist).
3.
Lock the electronics unit to the quadrupole head using the tightening
knobs (Figure 3-1) on the side of the electronics until it is SECURELY
hand-tightened. NEVER TWIST THE ELECTRONICS. Always insert it
straight in and pull it straight back.
TIGHTENING
KNOB
1
FOCUS (LENS 1)
2
REPELLER / CIS (LENS 2)
3
FILAMENT (-)
4
DC (+)
5
HEAD ID
KEY
10
1
9
2
8
3
7
4
6 5
6
EXIT LENS
7
DC (-)
8
FILAMENT (+)
9
MULTIPLIER
10 SOURCE
TIGHTENING
KNOB