Electronics Unit Maintenance and Troubleshooting
This section includes information on:
•
Maintaining the analyzer head.
•
Changing the Filament.
•
Disassembling the Source.
•
Cleaning the Source.
•
Reassembling the Source.
•
Cleaning the electron detector.
•
Troubleshooting the electronics unit.
The rest of the electronics unit should require no routine maintenance
during normal use.
Electronics and Sensor
Disconnect all power from the analyzer before starting these
procedures.
Touching any part of the analyzer head, Source, or Mass Filter
with your fingers will leave dirt and oil on the parts resulting in
contamination of the head. Use clean plastic/latex gloves when
handling components.
Before connecting the 24 VDC power to the electronics, ensure
that the electronics are connected to the sensor. Otherwise, damage
to the electronics may occur.
Always remove 24 VDC power from the AMETEK electronics
before removing the electronics from the sensor.