Replacing the Control Board

1.
2.
3.
4.
5.
Control
Board
Figure 5-3.
Control and Scan
boards.
The Control board is located behind the front heat sink and LED panel
on the left side (see Figure 5-1). It is the shorter of the two boards
attached to the LED panel.
To remove the Control board:
Remove the I/O board.
Remove the 1/4" screws from the front top and bottom casing of the
unit and the 3/16" screw from the right side (Figure 5-1).
Loosen or remove the screws at the top and bottom of the option plate
to allow easier access to the removal of the Control board.
Grab the heat sink and pull on it to remove the assembly with the
Control and Scan boards. It is connected to the unit by four
connectors.
Once the assembly has been removed from the unit casing, remove
the upper left and bottom right 3/8" screws with internal tooth lock
washer from the heat sink. See Figure 5-3.
Screw for
screw for
Control Board
control Board
Scan Board
scan Board
LEDs
lEds
Screw for
screw for
Scan Board
scan Board
Screws for
Heat Sink
Front Heat Sink
Front Heat sink
Board Replacement | 5-5