Appendix A – DCU Parameters Settings
DCU Parameters for High Vacuum System with an Oil Vane Pump and the std
normally open vent valve. No Turbo sealing gas valve.
794 Parameter Set
0 Lock Out
1 Open Select 1 allows Parameter Changes
010 Pumping System
0 Off (Default)
1 ON Select 1
012 Enable Venting
0 No (Default)
1 Yes Select 1
019 Cfg DO2
1 No Error (Default)
024 Cfg DO1
0 Rot Switch Speed Obtained (Default)
035 Cfg Acc A1|
0 Fan (Default)
036 Cfg Acc B1Vent Vlv (Default)
1 Vent Valve Normally Closed
8 Power FailureVent Valve Select 8, Normally Open Vent Valve
701 Spd Sw Pt
80% (Default)
97% Select 97
794 Parameter Set
0 Lock Out Select 0 to lock Parameters
1 Open
DCU Parameter Settings | A-1