OX-TRAN Model 2/28 Operator's Manual

Chapter 3: Preparing for a Test

This chapter provides information on how to prepare for a permeation test.
Read this chapter to learn about:

Testing Basics

How Barrier Properties Affect Testing

Developing Test Methods for Films
Developing a Temperature Profile
Testing Basics
A permeation test is comprised of a series of discrete actions or events. Some of these steps require human
thought, action or intervention. Some of the steps are performed by the instrument and will be discussed
elsewhere. The most important part of the process is the planning and preparation that occurs before the
actual testing begins. Inadequate planning or preparation may result in an inaccurate measurement. This
chapter contains information on the factors that should be considered when planning for a permeation test.
How Barrier Properties Affect Testing
A good understanding of the barrier material properties for the samples to be tested is very helpful when
preparing for a permeation test. The transmission rate of the sample and the conditions at which it will be
tested may require changes to the test methodology.
Any sample with a transmission rate within the range of the instrument (as specified in Chapter 8) can be
tested. Additional consideration is advised when setting up tests for samples that have a transmission rate
within 25% of the upper or lower end of the instruments test range.
Use caution when testing at the upper end of the instruments test range, an excessively high transmission
rate may over-range the oxygen sensor. Persistent or repeated over-range conditions will shorten the
lifespan of the Oxygen sensor. Common poor barrier materials include polyethylene, polycarbonate and
acrylic.
The possibility of an over-range condition can be reduced by; lowering the oxygen concentration of the Test
Gas, using a mask to reduce the sample area when testing film samples or increasing the Carrier Gas flow
rate. Using any or all of these methods to reduce the amount of oxygen the Oxygen Sensor is exposed to
will also increase the life of the sensor.

Testing Mixed Barriers

MOCON, Inc.
Revision J
Preparing for a Test
3-1