CARESCAPE ONE User Manual DRAFT 26 October 2017
You can select the patient in the host monitor to continue the host monitor
patient case. This option will only appear if a patient is admitted to the host
monitor. It is shown as the host monitor bed name plus the patient name and
MRN number.
You can merge the CARESCAPE ONE patient case with the host monitor patient
case. This option is shown as the host monitor bed name + CS ONE.
You can select New Patient to discharge the patient from the CARESCAPE ONE,
deleting all patient data and history, and start monitoring a new patient/case
at the host monitor
4. If needed, reconnect any parameters that were disconnected during transport.

Transport transfer settings

If a CARESCAPE ONE is taken to another unit and connected to a different host monitor
as an acquisition device, the CARESCAPE ONE uses the receiving host monitor's
care unit settings. For example, if the CARESCAPE ONE is disconnected from a host
monitor in the ICU, the care unit settings used during transport are those from the ICU
host monitor. If that same CARESCAPE ONE is then taken to the OR and connected
to a different host monitor there, upon connection the CARESCAPE ONE will use the
OR host monitor's care unit settings. Any changes made during transport when the
CARESCAPE ONE is not docked to a host monitor are not saved when the CARESCAPE
ONE is connected to another host monitor.
There are limits on the amount of data that the CARESCAPE ONE provides to a different
host monitor upon connection after transport from another host monitor. The
transferred history is limited to 36 ECG and arrhythmia related events, 10 ST events,
and 24 hours of trended data excluding high resolution trends.
For more information, see the Transport settings section in this chapter, and the About
continuing monitoring section in the Starting and ending monitoring chapter.

Transport settings

The following settings are transferred between the CARESCAPE ONE and the host
monitor when the CARESCAPE ONE is connected to the host monitor.
The listed settings are transferred when patient data is transferred; either when a
CARESCAPE ONE with an active patient case is connected to an empty host monitor
with no patient case, when a patient case from a CARESCAPE ONE is loaded to the
host monitor, or when an active CARESCAPE ONE patient case is merged with an
existing host monitor patient case.
If you make changes to any of the listed settings on the host monitor, the changes
are synchronized with the CARESCAPE ONE as long as the CARESCAPE ONE remains
connected to the host monitor.