CARESCAPE ONE User Manual DRAFT 26 October 2017
Default settings

NIBP default settings

Select Monitor Setup > Main Setup > Parameter Setup > Page 1 > NIBP > Setup > Page
1 or Page 2 to adjust these settings for the current patient. The following table lists the
factory default settings.
Setting
Description
NIBP Auto
Enable automatic
measurements.
Cycle Time
Select the
amount of time
between automatic
measurements.
Completed NIBP
Select the tone
Volume
volume that sounds
when an NIBP
measurement is
complete.
Display Format
Display the values
display format.
Auto Initial
Enable automatic
Inflate
selection of initial
cuff inflation
pressure.
Init. Pressure
Select the adult
mmHg
cuff initial inflation
pressure.
Select the child
cuff initial inflation
pressure.
Select the infant
cuff initial inflation
pressure.
Cuff Size
Select the cuff size.
Select Monitor Setup > Main Setup > Parameter Setup > Page 1 > NIBP > Alarms to adjust
these settings for the current patient. The following table lists the factory default settings.
358
ED
ICU
5 min
15 min
Sys/Dia (Mean)
(Not Selected)
(Adolescent profile:
Adult; Pediatric profile:
Child; Infant profile:
Infant)
CARESCAPE ONE
NICU
OR
Disabled
10 min
5 min
3
Enabled
135
125
100
Infant
(Not Selected)
(Adolescent profile:
Adult; Pediatric profile:
Child; Infant profile:
Infant)
2094480-001
PACU