Messages Related To Spo 2 Measurement With Carescape One - GE CARESCAPE ONE User Manual

Hide thumbs Also See for CARESCAPE ONE:
Table of Contents

Advertisement

CARESCAPE ONE User Manual DRAFT 26 October 2017
Messages
Message
Lead RL-LL failed
Measurement off
OFF
Relearning
Resp (Imped.)
measurement paused
Measurement paused
RR (Imped.) high / RR
(Imped.) low
Messages related to SpO
CARESCAPE ONE
For information regarding alarm priorities and escalation times, see the supplemental
information provided.
Make sure you are familiar with the generic layout of the screen. This will help you identify
where on screen the following messages appear. The message location is indicated with the
following abbreviations:
al. area = alarm area
param. = parameter window
wavef. = waveform area
214
Location
Possible causes
param.,
Lead RL-LL has been selected for
wavef.
measurement, but the RL or LL
electrode is off or is not adhered
to the patient well enough to get
a sufficient signal for respiration
monitoring.
wavef.
ECG leads not connected to the
patient.
param.
Measurement has been turned
off from the setup menu.
param.,
The patient's breathing pattern
wavef.
is being relearned because
the measurement has just
been turned on, the user has
activated relearning, a lead has
been changed, or a lead has
been reconnected after a lead
fail situation.
al. area
This alarm is triggered when
apnea alarm is enabled and
wavef.
respiration rate has been invalid
for the last 18 seconds in the
NICU software package or for
the last 50 seconds in other
software packages due to:
A lead fail.
Ongoing automatic
relearning.
Impedance respiration
malfunction.
al. area
Measurement values are equal
to or outside the alarm limits.
CARESCAPE ONE
Suggested actions
Check the electrodes and
their connections.
You may need to prep the skin
and replace the electrode.
Connect the ECG leads to
the patient to start the
impedance respiration
measurement.
Turn on the measurement
from the setup menu.
Wait until the message
disappears.
Check the patient status.
Check the electrodes and
their connections.
Acknowledge the alarm.
Check the patient status.
Adjust alarm limits if
necessary.
measurement with
2
2094480-001