Transport use

Preparing to transport a patient

1. Place the host monitor into standby mode. Refer to the host monitor user manual
for instructions.
2. Add parameters needed for transport to the CARESCAPE ONE, or remove those
not needed for transport.
3. Ensure that the parameters you need to monitor during transport are selected
to display on the CARESCAPE ONE screen. Refer to the chapter "Setting up the
monitor before use" for more information on selecting parameters to the screen.
4. Ensure that there is a charged battery installed in the CARESCAPE ONE.
5. Disconnect the CARESCAPE ONE from the CARESCAPE Dock F0.
6. Place the CARESCAPE ONE in the transport location. Use a Mini Dock and/or
a Parameter Dock 1 or Parameter Dock 5 to secure the CARESCAPE ONE and
CARESCAPE Parameters during transport.
WARNING
NOTE
NOTE
NOTE
7. Transport the patient.

Returning a patient from transport

Follow these steps to return the patient to the same host monitor after transport,
retaining the patient data.
1. Return the patient to the room.
2. Remove the CARESCAPE ONE from the Mini Dock.
3. Connect the CARESCAPE ONE to the CARESCAPE Dock F0 at the host monitor.
Patient monitoring will automatically resume on the host monitor if the patient
case matches. Patient data acquired on the CARESCAPE ONE during transport
will be available on the host monitor.
If the patient case does not match between the host monitor and the CARESCAPE
ONE, then the Select Patient and Data window will open on the host monitor.
There are four possible options:
You can select the patient in the CARESCAPE ONE to continue the CARESCAPE
ONE patient case. This option is shown as CS ONE, along with the patient name
and MRN number.
88
Do not place the CARESCAPE ONE in a patient bed.
Make sure the cables do not drag on the floor or interfere
with the movement or operation of the transport bed.
The cables must be placed so that they do not entangle
infusion pump lines or become a hazard to the patient.
To maintain visibility of the CARESCAPE ONE display, the
bedrail mount should be mounted on a rail or headboard
that provides support to the back of the mount. The
bedrail mount should be used on stationary bedrails or
headboards only.
The pole mount is intended for use on a vertical pole only,
and should not be used on a horizontal bedrail.
CARESCAPE ONE
2094480-001