CARESCAPE ONE User Manual DRAFT 26 October 2017

Supply mains interruption

If the supply mains to the equipment is interrupted for less than 2 hours and the
equipment shuts down, the monitor keeps the trend data and the latest user-made
settings. After 2 hours, all patient information and trend data is lost, and the monitor
returns to the user default settings. For more information, see the service manual.

User interface indicators

The following indicators appear in the software user interface.
Alarm volume adjustment for high and medium priority. Can also be used to adjust low
priority alarm volume when high, medium, and low priority alarms are configured in
Care Unit Settings to Common for All.
Alarm volume adjustment for low priority.
Audio alarms off indicator.
Audio alarms paused indicator with countdown timer - Indicates all audio alarms are
paused and the amount of time remaining for the alarm pause period displays as a
countdown timer.
Alarms audio pause indicator. Indicates that alarm audio pause has been activated.
Acknowledge alarms indicator. Indicates that the alarm can be acknowledged by
touching the alarm message.
Low priority audio off alarm indicator.
General warning sign. Displays when the priority setting deviates from the
recommendation of international alarm safety standards.
Touch indicator.
Home indicator. Close all menus/applications displayed on the monitor.
Patient indicator.
CARESCAPE ONE battery is full indicator.
CARESCAPE ONE battery indicator (green). The higher the charge, the bigger the green
bar within the indicator.