1. Select the HR parameter window.
2. Select Setup > Page 2.
3. Select a parameter from the Primary HR Source list. The selection list will only
show active measurements and AUTO, IntelliRate, or ECG. Choices are:
Single HR: AUTO, IntelliRate, ECG, invasive pressure channel(s) labeled Art X,
Fem X, or UAC X where X = channel number, Pleth (SpO2).
Multiple HR: AUTO, IntelliRate, ECG.
For more information about AUTO and IntelliRate, see the ECG measurement
practicalities section in this chapter.

Selecting HR alarm range

With the Single HR Alarms setting only.
1. Select the HR parameter window.
2. Select Advanced > Page 2.
3. Select the heart rate alarm range from the HR Alarm Range list:
30 - 240 (set and disabled if the Primary HR Source is IntelliRate, AUTO, or
Pleth (SpO2).)
20 - 300

Showing another value in the HR parameter window

In addition to the primary HR, you can display another value in the HR parameter
window.
1. Select the HR parameter window.
2. Select Advanced > Page 2.
3. Select an option to display in the HR parameter window from the Show with
HR list:
ST
PVC
2nd HR
Off
2094480-001
This setting adjusts the primary heart rate source for all of the
hemodynamic parameters.
If the primary HR source is ECG, AUTO (ECG), or IntelliRate (ECG), the
secondary HR source displayed in the order: primary arterial source (Art X,
Fem X, or UAC X, where X = channel number), SpO2.
If the primary HR source is anything other than mentioned above, the
secondary HR source is always ECG.
CARESCAPE ONE
ECG
109