Messages Related To Various Technical Issues - GE CARESCAPE ONE User Manual

Hide thumbs Also See for CARESCAPE ONE:
Table of Contents

Advertisement

CARESCAPE ONE User Manual DRAFT 26 October 2017
Message
Move CO2 sensor to a
cooler location
Sensor overheating
Over range
Over scale
RR (CO2) high
RR (CO2) low
Sample line
disconnected
Line disconnected
Service CO2 sensor
Check sensor
Warming up
Zeroing

Messages related to various technical issues

The following table lists messages that are not directly related to any parameter or
measurement. They are mostly technical messages related to hardware, configuration,
and similar issues.
2094480-001
Location
Possible causes
al. area
The LoFlo Sidestream Module
temperature is too high.
param.
param.
The CO
value is outside the
2
measurable range and cannot
be measured
wavef.
The CO
signal exceeds the
2
waveform area.
al. area
Measurement values are equal
to or outside the alarm limits.
al. area
The sample line is disconnected.
param.
al. area
There is a problem with the LoFlo
Sidestream Module.
param.
param.
The LoFlo Sidestream Module is
warming up.
wavef.
Zeroing is in progress.
param.
CARESCAPE ONE
Messages
Suggested actions
Check the LoFlo Sidestream
Module placement and
ensure that it is not covered
by anything that might lead
to overheating.
Replace the LoFlo Sidestream
Module.
Contact qualified service
personnel.
Zero CO
.
2
Replace the CARESCAPE CO
Contact qualified service
personnel.
Check the patient status.
Select a larger scale for the
waveform.
Check the patient status.
Adjust alarm limits if
necessary.
Check the sample line
connection and reconnect
as needed, then perform
zeroing.
Replace the LoFlo Sidestream
Module.
Contact qualified service
personnel.
Wait until the warm-up is
completed and the message
disappears.
Wait until the zeroing is
completed and the message
disappears.
.
2
229