CARESCAPE ONE User Manual DRAFT 26 October 2017
Message
Move CO2 sensor to a
●
cooler location
Sensor overheating
●
Over range
●
Over scale
●
RR (CO2) high
●
RR (CO2) low
●
Sample line
●
disconnected
Line disconnected
●
Service CO2 sensor
●
Check sensor
●
Warming up
●
Zeroing
●
Messages related to various technical issues
The following table lists messages that are not directly related to any parameter or
measurement. They are mostly technical messages related to hardware, configuration,
and similar issues.
2094480-001
Location
Possible causes
al. area
The LoFlo Sidestream Module
●
temperature is too high.
param.
●
param.
●
The CO
value is outside the
2
measurable range and cannot
be measured
wavef.
The CO
signal exceeds the
●
2
waveform area.
al. area
Measurement values are equal
●
to or outside the alarm limits.
al. area
The sample line is disconnected.
●
param.
●
al. area
There is a problem with the LoFlo
●
Sidestream Module.
param.
●
param.
●
The LoFlo Sidestream Module is
warming up.
wavef.
Zeroing is in progress.
●
param.
●
CARESCAPE ONE
Messages
Suggested actions
Check the LoFlo Sidestream
●
Module placement and
ensure that it is not covered
by anything that might lead
to overheating.
Replace the LoFlo Sidestream
●
Module.
Contact qualified service
●
personnel.
Zero CO
.
●
2
Replace the CARESCAPE CO
●
Contact qualified service
●
personnel.
Check the patient status.
●
Select a larger scale for the
●
waveform.
Check the patient status.
●
Adjust alarm limits if
●
necessary.
Check the sample line
●
connection and reconnect
as needed, then perform
zeroing.
Replace the LoFlo Sidestream
●
Module.
Contact qualified service
●
personnel.
Wait until the warm-up is
●
completed and the message
disappears.
Wait until the zeroing is
●
completed and the message
disappears.
.
2