CARESCAPE ONE User Manual DRAFT 26 October 2017
Alarms
Power loss
duration
Up to 2 hours
Greater than 2
hours
Alarm data stored in Clinical logs
Access to the Clinical logs is a service-level function and it is password protected.
CARESCAPE ONE stores a record of patient-related alarms and information messages
as well as any adjustments to the alarm limits in the Clinical logs.
Stored alarm data during a power cycle or power loss
If the CARESCAPE ONE goes through a power cycle or a loss of power, the stored
alarm data in the Clinical log is not affected. The alarm data remains stored in the
Clinical log until the monitor automatically clears the oldest stored data to allow new
data to be stored.
86
Alarm setting status after a power loss
The alarm settings that are in effect before the power loss are restored
automatically.
The alarm settings revert back to the user default settings (startup
profile). You must reconfigure any patient-specific alarm settings.
CARESCAPE ONE
2094480-001