CARESCAPE ONE User Manual DRAFT 26 October 2017
SpO₂
SpO
compatibility limitations
2
For detailed information regarding CARESCAPE ONE, CARESCAPE Parameter, and
accessory compatibility, see the supplemental information provided.
SpO
safety precautions
2
SpO
warnings
2
WARNING
WARNING
WARNING
WARNING
WARNING
2094480-001
DEFIBRILLATOR PRECAUTIONS. Patient signal inputs labeled
with the CF and BF symbols with paddles are protected against
damage resulting from defibrillation voltages. To ensure
proper defibrillator protection, use only the recommended
cables and leadwires. Using other cables or leadwires may
result in damage to the equipment and compromise patient
and user safety.
The operator is responsible for checking the compatibility of
the pulse oximetry device, sensor, and patient cable prior
to use. Incompatible components can result in degraded
performance and/or device malfunction.
If the accuracy of any measurement does not seem
reasonable, first check the patient's vital signs, then check
for conditions that may cause inaccurate SpO
the problem is still not resolved, check the CARESCAPE ONE,
CARESCAPE SpO
, and sensor for proper functioning.
2
A pulse oximeter should not be used as an apnea monitor. A
pulse oximeter should be considered an early warning device.
As a trend toward patient deoxygenation is indicated, blood
samples should be analyzed by a laboratory CO-oximeter to
completely understand the patient's condition.
Check that the pulse oximetry waveform is physiological
in shape to ensure waveform quality and minimize noise
spikes caused by motion conditions. (Not applicable when
monitoring SpO
with Masimo SET technology.)
2
CARESCAPE ONE
11
readings. If
2
133