CARESCAPE ONE User Manual DRAFT 26 October 2017
Messages related to temperature measurement
For information regarding alarm priorities and escalation times, see the supplemental
information provided.
Make sure you are familiar with the generic layout of the screen. This will help you identify
where on screen the following messages appear. The message location is indicated with the
following abbreviations:
al. area = alarm area
●
param. = parameter window
●
wavef. = waveform area
●
Message
Calibration fail
●
T1 Calibration fail / T2
●
Calibration fail
Check device
●
Temperature
●
malfunction
Device overheating
●
Move temp device to
●
a cooler location
Identical Temp device
●
No sensor detected
●
T1 high / T1 low
●
T2 high / T2 low
●
T2–T1 high
●
Temp measurement
●
removed
2094480-001
Location
Possible causes
param.
●
Calibration of a channel failed.
al. area
●
param.
●
Hardware or software fault in
CARESCAPE Temperature:
al. area
●
Cable fault
●
Data out of expected range
●
param.
●
CARESCAPE Temperature
surface temperature is too high.
al. area
●
al. area
There are two CARESCAPE
●
Temperatures.
param.
●
No probe detected, or the probe
is not compatible. Also when the
temperature probe check fails.
al. area
Measurement values are equal
●
to or outside the alarm limits.
al. area
Measured delta value is equal to
●
or outside the alarm limits.
CARESCAPE Temperature
al. area
●
has been removed or active
temperature channel becomes
inactive.
CARESCAPE ONE
Messages
Suggested actions
Check connections.
●
If the problem persists,
●
contact qualified service
personnel.
Reconnect the CARESCAPE
●
Temperature.
Replace the CARESCAPE
●
Temperature.
If the problem persists,
●
contact qualified service
personnel.
Check the CARESCAPE
●
Temperature placement and
ensure that it is not covered
by anything that might lead
to overheating.
Replace the CARESCAPE
●
Temperature.
Remove one CARESCAPE
●
Temperature.
Check the probe and
●
connections.
Ensure that you are using a
●
compatible probe.
Check the patient status.
●
Adjust alarm limits if
●
necessary.
Check the patient status.
●
Adjust alarm limits if
●
necessary.
Check all connections and
●
reconnect as required.
225