Co 2 Standards Compliance; Co 2 Performance Specifications - GE CARESCAPE ONE User Manual

Hide thumbs Also See for CARESCAPE ONE:
Table of Contents

Advertisement

CARESCAPE ONE User Manual DRAFT 26 October 2017
Temperature probe types supported
Temperature test measurement cycle
Reusable and disposable temperature
probe support
Replaceable sensor interface cable
CO
standards compliance
2
The system with CARESCAPE CO
The CARESCAPE CO
system are TYPE BF DEFIBRILLATION-PROOF APPLIED PARTs per ISO 80601-2-55
Clause 201.4.6.
CO
performance specifications
2
CO
range
2
CO
accuracy
2
Resolution
Stability
awRR (airway respiratory rate) range
awRR (airway respiratory rate) accuracy
Warm-up time
Sample flow rate
Total system response time
Total system rise time
CO
sweep speed options
2
2094480-001
Series 400
Every minute
Sensor interface cables provide reusable or disposable
temperature probe interfaces.
Tool removable sensor interface cable
complies with ISO 80601-2-55:2011-12.
2
accessories that are intended to be connected with the breathing
2
0-19.7%
After two minutes warm-up:
For values between 0 and 40 mmHg (0 and 5.3 kPa): ±2.0
mmHg (±0.29 kPa)
For values from 41–70 mmHg (5.4–9.3 kPa): ±5% of reading
For values from 71–100 mmHg (9.4–13.3 kPa) ±8% of reading
For values from 101–150 mmHg (13.4–20 kPa): ±10% of reading
At respiration rates above 80 rpm, all ranges are ±12% of reading.
The specifications are valid for gas mixtures of CO
dry gas at 760 mmHg (101.3 kPa) within specified operating
temperature range.
Numeric: 1.0 mmHg (0.1 kPa)
Wave: 0.1 mmHg (0.01 kPa)
Short-term drift: ±0.8 mmHg (0.13 kPa) over six hours
Long-term drift: Accuracy specification is maintained over a
120-hour period
2-150 rpm
± 1 rpm
2 minutes with CO
sensor attached for full accuracy specification
2
50 ±10 ml/minute
3 seconds for on-airway adapter kits
(Additional 30 ms for sidestream sampling cannulas) (Additional 2
seconds for extension line and dehumidification tubing)
200 ms for on-airway adapter kits
(Additional 30 ms for sidestream sampling cannulas) (Additional 80
ms for extension line and dehumidification tubing)
0.625, 6.25, 12.5, 25, and 50 mm/s
CARESCAPE ONE
Parameter specifications
, balance N
,
2
2
321