CARESCAPE ONE User Manual DRAFT 26 October 2017
For information on materials used in accessories and their biocompatibility, refer
●
to the instructions for use in the accessory package.
The water trap is contained within the sample lines.
●
Place the airway adapter with all sampling ports upwards.
●
Always check the tightness of all connections.
●
Materials used in accessories are not toxic. Accessories do not contain natural
●
rubber latex. Also see the instructions for use in the accessory package.
The system automatically performs barometric pressure compensation to meet
●
the stated accuracy of the LoFlo Sidestream Module.
Nitrous oxide, elevated levels of oxygen, helium, xenon, halogenated hydrocarbons,
●
and barometric pressure can influence the CO
LoFlo.
If nitrous oxide is used, always use N
●
CO₂ measurement setup
Equipment connection with CARESCAPE CO
For intubated and non-intubated patients.
1. CARESCAPE ONE
2. CARESCAPE CO
3. Exhaust port
4. LoFlo Sidestream Module
5. Accessory port (for sample lines, nasal cannulas, etc.)
Preparing the setup for CARESCAPE CO
1. Connect the LoFlo Sidestream Module to CARESCAPE CO
2. Connect CARESCAPE CO
3. Connect the sample line to the LoFlo Sidestream Module.
4. Zero the LoFlo Sidestream Module.
2094480-001
O compensation.
2
2
to the CARESCAPE ONE.
2
CARESCAPE ONE
measurement with Respironics
2
2
2
.
2
CO₂
187