CARESCAPE ONE User Manual DRAFT 26 October 2017
SpO₂
SpO
functional testers
2
You can verify the functionality of pulse oximeter sensor and monitor with a functional
SpO
tester but you cannot evaluate their accuracy with such a device. For more
2
information, refer to the standard ISO 80601-2-61 Annex FF (Simulators, calibrators
and functional testers for pulse oximeter equipment).
Using the SpO₂ measurement
Changing the SpO
NOTE
1. Select the SpO
2. Select Setup > Page 1.
3. Select the scale from the Scale list:
AUTO: The scale is automatically selected according to the IrMod % (infrared
modulation percentage) received from the measurement source.
Other scale options are 2, 5, 10, 20, or 50 %.
Changing the SpO
NOTE
1. Select the SpO
2. Select Setup > Page 1.
3. Select the size from the Size list.
Selecting the SpO
NOTE
1. Select the SpO
2. Select Setup > Page 1.
3. Select a numeric value from the Hemo Sweep Speed list.
The smaller the value, the slower the sweep speed.
Selecting the SpO
The primary heart rate can be calculated from the ECG leads, SpO
or invasive pressure waveform.
NOTE
NOTE
144
waveform scale
2
GE TruSignal technology only.
parameter window.
2
waveform size
2
Masimo SET technology and Nellcor
technology only.
parameter window.
2
hemodynamic sweep speed
2
This setting adjusts the waveform speed for all of the
hemodynamic parameters.
parameter window.
2
as the primary heart rate source
2
This setting adjusts the primary heart rate source for all of the
hemodynamic parameters.
HR Alarms must be configured as Single to enable the SpO
as the primary heart rate source. This setting is in Care Unit
Settings and is password protected.
CARESCAPE ONE
sensors with OxiMax
measurement,
2
2
2094480-001