CARESCAPE ONE User Manual DRAFT 26 October 2017
2. Check that the monitor recognizes cable connections (activates the display) for all
the pressure channels used and the pressure values and appropriate waveforms
are displayed.

Invasive pressure measurement on the monitor screen

The invasive pressure channel labels are as follows. "X" indicates the channel number,
for example, Art 2. Throughout this document, when an invasive pressure label is
followed by an "X", the "X" represents the channel number.
With the exception of PA, more than one channel can have the same channel label, for
example, Art 1 and Art 2. Only one channel can be labeled PA.
Label
Art X
Fem X
FemV X
PA X
CVP X
LAP X
RAP X
ICP X
RVP X
UAC X
UVC X
P1 to P2

Using the invasive pressure measurement

Invasive pressure analog output

CARESCAPE ONE sets the first available arterial channel (Art X, FemX, UAC X) for
invasive pressure analog output.
If there are no arterial channels available, the first zeroed channel will be used.
If there are no zeroed channels available, a flat line will be the output of the IP
Analog output channel.

About zeroing the invasive pressure transducers

Prior to monitoring, zero transducers at the patient's phlebostatic axis. Zeroing the
pressure transducers is very important for accurate pressure measurements. To
avoid inaccurate measurements, you must zero the pressure transducers:
Before measuring invasive pressures.
Before initiating treatment changes reliant upon pressures data.
When using a new transducer or tubing.
After reconnecting the transducer cable to the CARESCAPE Pressure.
2094480-001
Description
Arterial pressure
Femoral arterial pressure
Femoral venous pressure
Pulmonary arterial pressure
Central venous pressure
Left atrial pressure
Right atrial pressure
Intracranial pressure
Right ventricular pressure
Umbilical arterial pressure
Umbilical venous pressure
Non-specific pressure channel labels
CARESCAPE ONE
Invasive pressures
167