Alarm Priorities And Escalation Times For Impedance Respiration - GE CARESCAPE ONE User Manual

Hide thumbs Also See for CARESCAPE ONE:
Table of Contents

Advertisement

CARESCAPE ONE User Manual DRAFT 26 October 2017
Alarm specifications
MESSAGE
NOTES
Trigeminy
V Brady
V Fib/V Tach
V Tach
VT>2
Alarm priorities and escalation times for impedance
respiration
The following table shows the messages with their priorities and escalation times. In the
table, the message location is indicated by the following: alarm = alarm area; param. =
parameter window; and wavef. = waveform area. In the following table, 0 s = no intended
delay. For more information on alarm messages, refer to the user information provided.
MESSAGE
NOTES
Apnea (Imped.)
OR sw package,
adjustable limit
seconds
PACU, ED, ICU,
Apnea (Imped.)
NICU sw
packages,
adjustable limit
seconds
Apnea (Imped.)
Alarm priority
set to info, low,
medium or high
Apnea
deactivated
Artifact
Cardiac artifact
278
LOCATION
PRIORITIES AND ESCALATION TIMES
info
alarm,
0s, according to priority setting: high, medium, low, off,
wavef.
info
alarm,
0s, according to priority setting: high, medium, low, off,
wavef.
info
alarm,
param.,
wavef.
alarm,
0s, according to priority setting: high, medium, low, info
param.,
The alarm priority is always high if the duration is over
wavef.
30 s, the set HR high limit is exceeded, and the HR is
over 180 in NICU sw package or over 150 in all other
sw packages.
alarm,
0s, according to priority setting: high, medium, low, off,
wavef.
info
LOCATION
PRIORITIES AND ESCALATION TIMES
info
alarm,
wavef.
alarm,
wavef.
alarm
X s, according to priority setting: high, medium, low,
info (X = adjustable limit)
param.
0 s
param.,
0 s
wavef.
alarm
CARESCAPE ONE
low
medium
low
medium
X + 40 s
X seconds
(X =
adjustable
limit)
X + 20 s
X seconds
(X =
adjustable
limit)
0 s
high
0 s
high
X +102 s
X + 40 s
2094480-001