Nibp Volume And Display Settings; Adjusting The Nibp Measurement Completion Tone Volume; Setting The Nibp Display Format; Nibp Alarms - GE CARESCAPE ONE User Manual

Hide thumbs Also See for CARESCAPE ONE:
Table of Contents

Advertisement

CARESCAPE ONE User Manual DRAFT 26 October 2017
Non-invasive blood pressure
NOTE
Selecting the initial NIBP cuff inflation pressure
You can determine the cuff inflation pressure automatically based on the cuff size.
1. Select the NIBP parameter window > Page 2.
2. Select Auto Initial Inflate.
Setting the target NIBP inflation pressure
You can manually change the target inflation pressure for the first NIBP measurement.
1. Select the NIBP parameter window > Page 2.
2. Check that Auto Initial Inflate is not selected.
3. Select the Init. Pressure mmHg (or Init. Pressure kPa) value with arrows.

NIBP volume and display settings

Adjusting the NIBP measurement completion tone volume

1. Select the NIBP parameter window > Page 2.
2. Set the Completed NIBP Volume.
The lower the value, the softer the tone.

Setting the NIBP display format

1. Select the NIBP parameter window > Page 2.
2. Select the format from the Display Format list:
Sys/Dia (Mean): All values are shown, but the sys/dia values are shown in a
bigger font.
(Mean) Sys/Dia: All values are shown, but the mean value is shown in a bigger
font.

NIBP alarms

Setting NIBP alarms

NOTE
1. Select the NIBP parameter window.
2. Select the Alarms tab.
3. Select Systolic (SYS), Mean (M), or Diastolic (DIA) pressure.
If the feature is not active, the alarm limits are greyed out. Select Alarm On to
set the alarms.
160
When the Auto Initial Inflate setting is enabled, the initial cuff
inflation pressures are dependent on the selected cuff size.
The initial target pressure preset can be adjusted if you desire
a lower or higher initial target pressure.
The most recent saved alarm on/off settings apply to all cuff
sizes.