CARESCAPE ONE User Manual DRAFT 26 October 2017
Messages
Message
RVP 1 mean high /
●
RVP 1 mean low to
RVP 2 mean high /
RVP 2 mean low
UAC 1 disconnect to
●
UAC 2 disconnect
Disconnected
●
UAC 1 sys high / UAC
●
1 sys low to UAC 2 sys
high / UAC 2 sys low
UAC 1 mean high /
●
UAC 1 mean low to
UAC 2 mean high /
UAC 2 mean low
UAC 1 dia high / UAC
●
1 dia low to UAC 2 dia
high / UAC 2 dia low
UVC 1 mean high to
●
UVC 2 mean high /
UVC 1 mean low to
UVC 2 mean low
Zeroed
●
Zeroing
●
Zeroing failed
●
Zero ICP separately
●
224
Location
Possible causes
al. area
Measurement values are equal
●
to or outside the alarm limits.
al. area
No arterial invasive pressure is
●
detected.
param.
●
al. area
Measurement values are equal
●
to or outside the alarm limits.
al. area
Measurement values are equal
●
to or outside the alarm limits.
param.
●
Zeroing was successful.
param.
●
IP channel is currently being
zeroed.
al. area
Pulsating waveform detected.
●
Defective transducer
Offset is >150 mmHg.
al. area
The ICP channel must be zeroed
●
separately from all other invasive
pressures.
CARESCAPE ONE
Suggested actions
Check the patient status.
●
Adjust alarm limits if
●
necessary.
Check the patient status.
●
Check connections.
●
If pressure drops because of
●
zeroing, perform the zeroing
process.
Check the patient status.
●
Adjust alarm limits if
●
necessary.
Check the patient status.
●
Adjust alarm limits if
●
necessary.
No action required.
●
Message is automatically
removed after 10 seconds.
No action required.
●
Message is automatically
removed and replaced with the
zeroing results after completion.
Open the transducer to room
●
air and zero the channel.
Replace the transducer, open
●
it to room air, and zero the
channel.
Zero the channel using the
●
Zero option found under the
ICP channel setup menu.
2094480-001