CARESCAPE ONE User Manual DRAFT 26 October 2017
Messages
Message
RVP 1 mean high /
RVP 1 mean low to
RVP 2 mean high /
RVP 2 mean low
UAC 1 disconnect to
UAC 2 disconnect
Disconnected
UAC 1 sys high / UAC
1 sys low to UAC 2 sys
high / UAC 2 sys low
UAC 1 mean high /
UAC 1 mean low to
UAC 2 mean high /
UAC 2 mean low
UAC 1 dia high / UAC
1 dia low to UAC 2 dia
high / UAC 2 dia low
UVC 1 mean high to
UVC 2 mean high /
UVC 1 mean low to
UVC 2 mean low
Zeroed
Zeroing
Zeroing failed
Zero ICP separately
224
Location
Possible causes
al. area
Measurement values are equal
to or outside the alarm limits.
al. area
No arterial invasive pressure is
detected.
param.
al. area
Measurement values are equal
to or outside the alarm limits.
al. area
Measurement values are equal
to or outside the alarm limits.
param.
Zeroing was successful.
param.
IP channel is currently being
zeroed.
al. area
Pulsating waveform detected.
Defective transducer
Offset is >150 mmHg.
al. area
The ICP channel must be zeroed
separately from all other invasive
pressures.
CARESCAPE ONE
Suggested actions
Check the patient status.
Adjust alarm limits if
necessary.
Check the patient status.
Check connections.
If pressure drops because of
zeroing, perform the zeroing
process.
Check the patient status.
Adjust alarm limits if
necessary.
Check the patient status.
Adjust alarm limits if
necessary.
No action required.
Message is automatically
removed after 10 seconds.
No action required.
Message is automatically
removed and replaced with the
zeroing results after completion.
Open the transducer to room
air and zero the channel.
Replace the transducer, open
it to room air, and zero the
channel.
Zero the channel using the
Zero option found under the
ICP channel setup menu.
2094480-001