Selecting Co 2 Average; Selecting The Fio 2 Level; Selecting The N 2 O Level; Selecting The Apnea Alarm Limit - GE CARESCAPE ONE User Manual

Hide thumbs Also See for CARESCAPE ONE:
Table of Contents

Advertisement

CARESCAPE ONE User Manual DRAFT 26 October 2017
CO₂
Selecting CO
2
You can select a time window that is used to select the displayed CO
highest peak EtCO
displayed.
1. Select the CO
2. Select Setup.
3. Select an option from the CO2 Average list.
Selecting the FiO
The presence of a large concentration of oxygen causes the CO
than the actual value. Use this option to compensate for the presence of O
1. Select the CO
2. Select Setup.
3. Select an option from the FiO2 level list: 21-40 or 40-100 percent.
Selecting the N
The presence of N
Use this option to compensate for the presence of N
1. Select the CO
2. Select Setup.
3. Select an option from the N2O level list: 0-40 or 40-80 percent.

Selecting the apnea alarm limit

The apnea alarm can be enabled to select a time delay before an apnea alarm is
generated. The delay is defined with the Apnea limit seconds setting. If anything
other than the default (20 seconds) is selected, the number of seconds selected for the
delay displays in the parameter window.
1. Select the CO
2. Select Alarms > Page 2.
3. Select an option from the Apnea limit seconds list.

Zeroing the LoFlo Sidestream Module

When the message CO2 zero required appears in the alarm messages area, or the
message Zero required appears in the CO
Module needs to be zeroed.
NOTE
You should also zero the LoFlo Sidestream Module after connecting or reconnecting
a sample line.
1. Disconnect the patient from CO
2. Connect the LoFlo Sidestream Module and, if necessary, wait until the sensor
warm-up message disappears.
190
average
value within the time window is selected as the CO
2
parameter window.
2
level
2
parameter window.
2
O level
2
O causes the CO
2
parameter window.
2
parameter window.
2
For best results, allow five minutes for module warm-up
before starting the zero.
CARESCAPE ONE
value to appear higher than the actual value.
2
O.
2
parameter window, the LoFlo Sidestream
2
monitoring.
2
value. The
2
value
2
level to appear lower
2
.
2
2094480-001