CARESCAPE ONE User Manual DRAFT 26 October 2017
Electromagnetic compatibility
Parameter
Essential performance
ECG analog output
No errors in ECG waveform amplitude greater than ± 10%.
●
Invasive blood
No errors in invasive blood pressure analog output greater than 1 V/100 mmHg ±
●
pressure analog
1.5% ± 1.0 mmHg or ± 0.5% ± 2 mmHg, whichever is greater.
output
Electromagnetic compatibility safety
precautions
Electromagnetic compatibility warnings
WARNING
WARNING
WARNING
Electromagnetic compatibility cautions
CAUTION
CAUTION
Electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an
environment.
388
The use of accessories, transducers and cables other
than those specified may result in increased emissions or
decreased immunity performance of the equipment or
system.
Other equipment may interfere with the system, even if that
other equipment complies with CISPR emission requirements.
The device/system should not be used adjacent to, or stacked
with, other equipment. Consult qualified personnel regarding
device/system configuration.
Use of known RF sources, such as cell/portable phones,
or other radio frequency (RF) emitting equipment near the
system may cause unexpected or adverse operation of
this device/system. Consult qualified personnel regarding
device/system configuration.
Changes or modifications to this device/system not expressly
approved by GE may cause EMC issues with this or other
equipment.
CARESCAPE ONE
2094480-001