Default settings

Current patient settings

Current patient settings are non-password protected settings that are adjusted to
meet the needs of the current patient. The current patient settings are not saved to
the monitor's permanent memory, but most current patient settings are configured to
persist over a warm start (CARESCAPE ONE is shut down for less than 2 hours).

Configuring care unit settings

To access and customize the care unit settings:
1. Select Monitor Setup > Defaults & Service > Default Setup.
2. Enter the Username: clinical.
3. Enter the Password: Change Me.
This default password should be changed.
4. Select Care Unit Settings.
5. Change the settings as needed in Alarms, Parameter Alarms, Admit/Discharge,
Units, and Parameters. All changes are automatically saved and applied.
To revert to the factory default care unit settings, end the case/discharge the
patient and select Factory Default > Yes.

Alarm settings

Select Care Unit Settings > Alarms to define the alarm settings for the care unit. The
following table lists the factory default settings.
Setting
Description
Show Alarm Limits Enable the
current alarm
limits to display
in parameter
window.
Latching Alarms
Select the level
of alarm priority
to sound until
the event is
acknowledged.
Alarm Light
Enable the alarm
light to be turned
off (0% (Off)
Allowed).
Audio Pause Time
Select the number
(min)
of minutes to
pause audible
alarms.
326
ED
ICU
NICU
Enabled
High
Disabled
CARESCAPE ONE
OR
PACU
None
Enabled
Disabled
2
2094480-001

Advertisement