CARESCAPE ONE User Manual DRAFT 26 October 2017
2. Select the Audible & Visual tab.
3. Select the Audio Pause tab.
4. Select Activate All Audible Alarms.
Technical alarms' deactivation with the pause audio key
Certain technical alarms can be deactivated with the pause audio key. The Alarm
Deactivation setting is configured in the Care Unit Settings and it is password
protected.
For more information, see the supplemental information provided.
Technical alarms for which the deactivation with the pause audio key can be allowed
are:
ECG Leads Off
●
Arterial Disconnect (Art, Fem, UAC)
●
SpO2 Probe Off
●
Apnea alarms' deactivation with the pause audio key
Apnea alarms can be deactivated with the pause audio key if the Allow alarm
deactivation with the Audio Pause key for: setting Apnea (CO2/Imped.) is enabled in
the Care Unit Settings. This setting is password protected.
For more information, see the supplemental information provided.
Breakthrough alarms
The following alarms will break through when escalated to or activated at high priority
alarm condition regardless of the 2 to 5 minute audible alarm pause: Asystole, V Fib /
V Tach, V Tach (in all software packages), and Brady (in the NICU software package
only).
NOTE
Latched alarms
When alarms are latched, the audible alarm and visual message remain after the
alarm condition no longer exists. The audible alarm can be paused with the pause
audio key, and this also clears the alarm message from the screen. Alarms can be
configured to latch for high priority alarms only, all alarm priorities, or none. Only
physiological alarms can be latched. The Latching Alarms setting is configured in the
Care Unit Settings and it is password protected.
For more information, see the supplemental information provided.
Alarm settings after a power loss
If the CARESCAPE ONE loses power, the amount of time without power affects whether
or not you need to reset the alarm settings.
2094480-001
If Alarms Setup > Audible & Visual > Audio Pause > Pause
All Audio for 5 min is selected before an alarm is triggered,
no alarms break through the alarm pause. The pause audio
behavior is configured in Care Unit Settings and the setting
is password protected.
CARESCAPE ONE
Alarms
85