Alarm priorities and escalation times for temperature

The following table shows the messages with their priorities and escalation times. In the
table, the message location is indicated by the following: alarm = alarm area; param. =
parameter window; and wavef. = waveform area. In the following table, 0 s = no intended
delay. For more information on alarm messages, refer to the user information provided.
MESSAGE
NOTES
Calibration fail
Check device
Device
overheating
Identical Temp
device
Move Temp
device to a cooler
location
No sensor
detected
T1 calibration fail
/ T2 calibration
fail
T1 high / T1 low
T2 high / T2 low
T1 high / T1 low
Alarm priority set
to Escalating.
T2 high / T2 low
T2-T1 high
T2-T1 high
Alarm priority set
to Escalating.
Temp device
failure: Call
service
Temp
measurement
removed
2094480-001
LOCATION
PRIORITIES AND ESCALATION TIMES
info
param.
0 s
param.
0 s
LOCATION
PRIORITIES AND ESCALATION TIMES
info
param.
0 s
param.
0 s
param.
0 s
alarm
alarm
param.
0 s
alarm
60 s
alarm
alarm
alarm
60 s
alarm
alarm
alarm
0 s if no
patient
case/ no
admitted
patient
CARESCAPE ONE
Alarm specifications
low
medium
low
medium
0 s
0 s
0 s
60 s
60 s
60 s
120 s
60 s
60 s
60 s
120 s
0 s
0 s if active
patient
case/
admitted
patient
high
high
291