CARESCAPE ONE User Manual DRAFT 26 October 2017
Setting
P1-P2 Mean High
Limit/Low Limit
P1-P2 Mean Alarm On
Select Ventricular & Atrial tab to adjust the invasive pressures ventricular and atrial alarm
profile settings. The following table lists the factory default settings.
Setting
RAP Mean High
Limit/Low Limit
RAP Mean Alarm On
RVP Mean High
Limit/Low Limit
RVP Mean Alarm On
LAP Mean High
Limit/Low Limit
LAP Mean Alarm On

Invasive pressures settings for labels

Selecting a new channel label for a channel will apply the following label default settings
to the current patient. Once a new label is selected, the channel's default settings can be
changed for the current patient and also saved as the channel's profile setting defaults.
When any channel label other than the default channel label is selected:
Filter Hz: 12
When Art is selected as the channel label:
Catheter Disconnect: Enabled
Display Format: Sys/Dia (Mean)
Show Pulse Rate: Disabled
Primary Arterial Source: AUTO
Smart BP: Enabled
Scale mmHg (waveform):
2094480-001
ED
ICU
15/0
(Infant profile: 10/0)
Enabled
Enabled
(Neuro and
Trauma
profiles:
Disabled)
ED
ICU
15/0
(Pediatric and Infant
profiles: 10/0)
(US only: 15/-5)
(US only: ICU software
package Pediatric and Infant
profiles: 15/0)
35/5
(Pediatric and Infant
profiles: 25/5)
20/5
(Pediatric and Infant
profiles: 15/5)
CARESCAPE ONE
NICU
OR
5/0
(Infant profile: 10/0; Perfusion
profile: 15/-5)
Enabled
NICU
OR
5/0
15/0
(Pediatric and
Infant profiles:
10/0)
Enabled
25/5
(Pediatric and Infant
profiles: 25/5)
Enabled
10/2
(Pediatric and Infant
profiles: 15/5)
Enabled
Default settings
PACU
15/0
PACU
15/0
(US only:
15/-5)
(Pediatric and
Infant profiles:
10/0)
35/5
20/5
337