CARESCAPE ONE User Manual DRAFT 26 October 2017
NIBP total cycle time
Automatic cuff deflation conditions
Invasive pressure standards compliance
The system with CARESCAPE Pressure complies with IEC 60601-2-34:2011-05.
The CARESCAPE Pressure sensor interface cable and compatible transducers are TYPE
CF DEFIBRILLATION-PROOF APPLIED PARTs per IEC 60601-2-34 Clause 201.6.
Invasive pressure performance specifications
Number of invasive pressure channels
Invasive pressure measurement range
Invasive pressure measurement
accuracy
Invasive pressure display resolution
Invasive pressure units of measure
Invasive pressure zero balance range
Invasive pressure zero balance accuracy ±1 mmHg (±0.1 kPa)
Invasive pressure pulse rate range
Invasive pressure pulse rate accuracy
Invasive pressure pulse rate display
resolution
Invasive pressure transducer sites, site
name, and displayed values
2094480-001
20 to 40 seconds typical (depending on heart rate, pressure, and
motion artifact)
Power off
●
Adult and child cuff cycle time exceeding 125 seconds
●
Infant cuff cycle time exceeding 90 seconds
●
Adult and child cuff pressure exceeding 300 mmHg (40.0 kPa)
●
Infant cuff pressure exceeding 150 mmHg (20.0 kPa)
●
2
-98 to 349 mmHg (-13.1 to 46.5 kPa)
± 0.5 % ± 1.50 mmHg (excluding transducer)
± 4 % or ± 4 mmHg, which ever is greater (including transducer)
1 mmHg
mmHg or kPa
±150 mmHg (±20.0 kPa)
0 to 360 bpm
±2% or ±2 bpm, whichever is greater
1 bpm
Arterial (Art): Systolic, diastolic, mean, rate
●
Femoral (Fem): Systolic, diastolic, mean, rate
●
Femoral vein (FemV): Mean
●
Pulmonary artery (PA): Systolic, diastolic, mean
●
Central venous pressure (CVP): Mean
●
Intra-cranial pressure (ICP): Mean
●
Left atrial (LAP): Mean
●
Pressure channel (P1, P2): Mean
●
Right atrial (RAP): Mean
●
Right vein (RVP): Mean
●
Umbilical artery (UAC): Systolic, diastolic, mean, rate
●
Umbilical vein (UVC): Mean
●
CARESCAPE ONE
Parameter specifications