CARESCAPE ONE User Manual DRAFT 26 October 2017
WARNING
WARNING

EMC cautions

CAUTION

ESD safety precautions

To avoid electrostatic charges building up, it is recommended to store, maintain
and use the equipment at a relative humidity of 30% or greater.
To prevent applying a possible electrostatic charge to the ESD sensitive parts of
the equipment, touch the metallic frame of the component or a large metal object
located close to the equipment. When working with the equipment and specifically
when the ESD sensitive parts of the equipment may be touched, a grounded wrist
strap intended for use with ESD sensitive equipment should be worn. See the
documentation provided with the wrist straps for details of proper use. Floors
should be covered by ESD dissipative carpets or similar. Special ESD-protective
clothing, or optionally cotton-based clothing with natural fibres, should be used
when working with the component.
2094480-001
Use of this equipment adjacent to or stacked with other
equipment should be avoided because it could result in
improper operation. If such use is necessary, this equipment
and the other equipment should be observed to verify that
they are operating normally. Consult qualified personnel
regarding device/system configuration.
PARTIAL LOSS OF PARAMETERS. To prevent partial loss of
parameters, do not route patient cables along with the
AC mains power cord or ePort cable. Doing so may affect
acquisition of parameter data.
Use of known RF sources, such as cell/portable phones,
or other radio frequency (RF) emitting equipment near the
system may cause unexpected or adverse operation of
this device/system. Consult qualified personnel regarding
device/system configuration.