Stopping The Spo Measurement; How To Interpret The Spo₂ Values; Spo 2 Signal Strength; Spo 2 Waveform Quality - GE CARESCAPE ONE User Manual

Hide thumbs Also See for CARESCAPE ONE:
Table of Contents

Advertisement

CARESCAPE ONE User Manual DRAFT 26 October 2017
Stopping the SpO
1. Remove the SpO
2. Disconnect the sensor from CARESCAPE SpO
3. Disconnect the CARESCAPE SpO
4. Select
5. Discard single-use sensors.
For Masimo SET technology, always disconnect the RD SET, M-LNCS, LNCS, or
LNOP sensor from CARESCAPE SpO
the RD SET, M-LNCS, LNCS, or LNOP sensor after it has been repositioned.
Use only sensors and cables listed in the supplemental information provided.
How to interpret the SpO₂ values
SpO
signal strength
2
For GE TruSignal technology and Nellcor
strength is indicated with asterisks in the parameter window. The signal strength
indicator refers to the amplitude of the plethysmographic waveform, not the quality of
the waveform. Three asterisks indicate strong pulsation.
For Masimo SET technology, the signal strength indicator refers to Masimo's
proprietary measurement, Signal Identification and Quality (Signal IQ) indicator. The
Signal IQ provides an indicator of the assessment of the confidence in the displayed
SpO
value.
2
SpO
waveform quality
2
NOTE
Under normal conditions, the SpO
to) the arterial pressure waveform. The typical SpO
sensor location with the fewest noise spikes.
Normal waveform
If noise (artifact) is seen on the waveform because of poor sensor placement, the
photodetector may not be flush with the tissue. Check that the sensor is secured and
the tissue sample is not too thick. Pulse rate is determined from the SpO
which can be disrupted by hemodynamic pressure disturbances. Motion at the sensor
site is indicated by noise spikes in the normal waveform.
Abnormal waveform
2094480-001
measurement
2
sensor from the patient.
2
2
to acknowledge the SpO2 measurement removed alarm.
Not for Masimo SET technology.
waveform corresponds to (but is not proportional
2
CARESCAPE ONE
.
2
from the CARESCAPE ONE.
before repositioning the sensor. Reconnect
2
sensors with OxiMax
waveform can help the user find a
2
SpO₂
technology, signal
waveform,
2
149