Technical Alarm Delay Specifications - GE CARESCAPE ONE User Manual

Hide thumbs Also See for CARESCAPE ONE:
Table of Contents

Advertisement

CARESCAPE ONE User Manual DRAFT 26 October 2017
Alarm
Limit range
RR (CO2) high
4 to 60 breaths
per minute
NICU sw
package: 4 to
100 breaths per
minute
RR (CO2) low
4 to 60 breaths
per minute
NICU sw
package: 4 to
100 breaths per
minute

Technical alarm delay specifications

The following table lists the alarm delays for technical alarms. Any alarm delays that are less
than 5 s are given as <5 s.
Alarm
Analog output malfunction
Arrhythmia paused
Art X disconnect
Fem X disconnect
UAC X disconnect
Disconnected
Artifact
Calibration fail
T1 Calibration fail
T2 Calibration fail
Call service
Check CO2 adapter
Check device
CO2 device malfunction
Check device
ECG malfunction
Check device
Imped. malfunction
2094480-001
Limit increment
1 breath per
minute
1 breath per
minute
Parameter
ECG
Invasive pressures
ECG
Temperature
NIBP
CO
2
CO
2
ECG
Impedance respiration
CARESCAPE ONE
Alarm delay
21 s (average) (20 to 22 s range)
21 s (average) (20 to 22 s range)
Alarm delay
<5 s
21 s (average) (20 to 22 s range)
8 s
<5 s
<5 s
22 s (average) (22 s range)
42 s (average) (41 to 42 s range)
<5 s
<5 s
<5 s
Alarm specifications
269