O2 Supply Alarm Test - GE 9100c Technical Reference Manual

Anesthesia machine
Table of Contents

Advertisement

9100c
3.6 O
supply alarm test
2
3-10
M1207027
Reproduced from the electronic master in MATRIX
1. Set the system to On.
2. Set all flow controls to 3 L/min.
3. Disconnect the O
4. Make sure that:
The "No O
Pressure" alarm occurs.
2
The N
O (if equipped) and O
2
last.
Air (if equipped) flow continues.
The "No drive gas" alarm occurs as well on the ventilator if
the ventilator uses O
5. Turn all of the flow controls fully clockwise (zero flow).
6. Reconnect the O
2
004
MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English
pipeline supply or close the cylinder valve.
2
flows stop. The O
2
as the drive gas.
2
pipeline supply.
flow stops
2