Device Configuration

Select Language

Sets the system language to: English, French, German, Italian, Dutch, Portuguese, Russian, Spanish, or Swedish. Once a
new language is selected, the Responder 2000 cycles power.
Note: Maintenance mode text is always in English, regardless of the language selected.

Stored Data Management

Clear Event Logs

Clears all stored event logs. Always clear event logs after the Main board is replaced.

Clear Service Required Indicator

Clears the Service Required LED. Once a service fault is corrected, use the Clear Service Required menu item to turn
off the Service Required LED. If the LED does not clear, another fault is present.

Clear Settings Menu Password

Clears the Settings Menu password.

Clear All Stored Data

Clears all stored data, including:
Event Logs
Service Required Indicator
Settings Menu Password
All other Settings.

Send Event Logs to PC

Event logs may be downloaded to a PC and sent to Technical support to help troubleshoot a problem.
1.
Go to Service Mode (password HEART1).
2.
Select Stored Data Management > Send Event Logs To PC.
3.
On the PC, start the Windows-based ServiceLink2000 Application.
4.
Select the appropriate Com Port and click Start Com.
5.
Click Receive Events. A progress bar is displayed.
6.
When downloaded, save the events file to the desired location. A progress bar is displayed while the CSV file is
saved.
7.
The ServiceLink2000 program closes after the file is successfully saved.
2025653-048 Revision B