Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare

3 Route Survey

3.1 Step One — Sketch
Start preparing Route Survey by sketching a floor plan of the hospital or clinic which will receive
the equipment. Include all areas on the delivery route from outside the building to destination. See
Illustration
7-8.
Reference Numbers: Numbers in circles refer to Route Survey data. The Route Survey is a form
on which site data are listed (see
Illustration 7-8:
3.2 Step Two — Survey
Data concerning the intended delivery route are recorded on the Route Survey in the following
pages. Record all loading capacities, corridor widths, door openings, turning radii, flooring
materials, elevator sizes, obstructions and so on.
246
Section
3.2).
Direction 5160944-13-1EN, Revision 2
3 Route Survey