Default alarm limits
SpO
high Off, low 90%
2
PR high 160, low 40
NOTE: For each SpO
package for patient weight limits and accuracy.
Modules with CO
2
Sampling rate:
Maximum sampling line length:
Sampling delay:
Total system response time:
Warm-up time:
Autozeroing interval:
Automatic compensation for barometric pressure.
Gas values are measured in ATPD conditions (ambient temperature and pressure, dry). When CO2 is
displayed as a partial pressure (kPa, mmHg), the value can be alternatively shown as wet (BTPS, body
temperature and pressure saturated).
Non-disturbing gases are those with a maximum effect on the CO
effect is valid for specific concentrations shown in parentheses of the non-disturbing gas:
Ethanol C
Acetone (<0.1%)
Methane CH
Nitrogen N
water vapor (0 to 100%)
Dichlorofluoromethane (<1%)
Tetrafluoroethane (<1%)
Disturbing gases and their effect on the CO
reflect the effect of specific concentrations (shown in parentheses) of an individual disturbing gas and
should be combined when estimating the effect of gas mixtures:
Halothane (4%) increases < 0.3 vol%
Isoflurane(5%) increases < 0.4 vol%
Enflurane(5%) increases < 0.4 vol%
Desflurane(24%) increases < 1.2 vol%
Sevoflurane(6%) increases < 0.4 vol%
Helium (50%) decreases < 0.3 vol%
If O
compensation is not activated:
2
If O
compensation is activated:
2
If N
O compensation is not activated: N
2
If N
O compensation is activated:
2
4 The reported SpO
pulse rate accuracy is the product specification. SpO
2
yielded a PR error of less than 2 bpm (rms) over the whole measurement range.
1 Limits are adjustable:
2-34
1
:
accessory, refer to the instructions for use in the accessory
2
measurement, N-FC and N-FCREC
H
OH (<0.3%)
2
5
(<0.2%)
4
(0 to 100%)
2
OFF to 51% for SpO
50 to 100% for SpO
250 to 35 bpm for PR high
30 to 245 bpm for PR low
150 ±25 ml/min (sampling line 2 to 3 m, normal conditions)
6 m
2.1 s typical with a 3-m sampling line
2.4 seconds typical with a 3-m sampling line, including sampling
delay and rise time (typically 3.7 seconds with a 6-m sampling
line)
1 min for operation, 30 min for full specification
4, 15, 30 and 60 minutes after start-up, then every 60 minutes
reading at 5.0 vol-% CO
2
O
(40 to 95%) decreases < 0.3 vol%
2
O
(40 to 95%) error < 0.15 vol%
2
O (40%) decreases < 0.4 vol%
2
N
O (40 to 80%) error < 0.3 vol%
2
pulse rate accuracy tests with a simulator
2
high
2
low
2
reading at 5.0 vol% < 0.2 vol%. The
2
are shown below. Errors listed
2