Alarms Related To The Spo 2 And Pulse Rate Measurement - GE Portrait Hub01 User Manual

Mobile monitoring solution
Table of Contents

Advertisement

Alarm message
Hub:
Weak Resp
connection
Central Viewer:
Weak Resp
connection
Hub:
Resp connection lost
Connection lost
Central Viewer:
Resp connection lost
Hub:
Resp device
overheating
Device overheating
Central Viewer:
Resp device
overheating
Alarms related to the SpO
measurement
Alarm priorities are given in the section discussing alarm specifications.
Make sure that you are familiar with the generic layout of the Hub and Central Viewer
screens. This will help you identify where the following alarm messages appear. The alarm
location is indicated with the following abbreviations:
al. area = alarm area
data = data area
5815148-01
Location
Possible explanations
al. area
Connection between the Hub
and sensor is poor and there
was a data loss between the
Hub and the sensor.
Hub:
Connection between the Hub
and sensor was lost due to too
al. area
long a distance between them
data
(for instance, the patient has
walked away from the Hub).
Central
Viewer:
al. area
Hub:
Sensor battery surface
temperature is too high.
al. area
The maximum surface
data
temperature of the devices
Central
being worn by the patient (Hub,
Viewer:
sensor battery and wearable
sensors) will never exceed
al. area
43°C (The device will shut
down before the temperature
reaches 43°C). When the surface
temperature is approaching
43°C, an overheating alarm
is given, allowing the clinician
to take appropriate action to
ensure the time of exposure is
limited based on the condition
of the patient.
Portrait
Mobile Monitoring Solution
Suggested actions
Place the Hub closer to the
patient.
If the problem persists,
contact qualified service
personnel.
Check the distance between
the sensor and the Hub.
Make sure the patient
understands how far they
can move from the Hub
without losing the wireless
connection.
If the problem persists,
contact qualified service
personnel.
Check the sensor battery and
ensure that it is not covered
by anything that might lead
to overheating.
If the problem persists,
replace the sensor battery.
and pulse rate
2
Alarm messages
139

Advertisement

Table of Contents
loading

This manual is also suitable for:

Portrait sbt01