Installation Reminders; Average Installation Time; Installation Warnings - GE Vivid 7 Service Manual

Hide thumbs Also See for Vivid 7:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FC091194, R
IRECTION
EVISION
Section 3-2

Installation Reminders

3-2-1

Average Installation Time

Table 3-2
Description
Unpacking the scanner
Install Scanner wo/options
DICOM Network Configuration
3-2-2

Installation Warnings

WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE
DANGER
AC GROUND LINE (I.E., METER'S GROUND SWITCH IS OPEN), DON'T TOUCH
THE UNIT!
If the unit is very cold or hot, do not turn on its power until it has had a chance to acclimate to
CAUTION
its operating environment.
Table 3-3
DON'T STORE AT
THESE
TEMPERATURES
o
C
-40
-35
-30
-25
o
F
-40
-31
-22
-13
Hrs
20
18
16
14
TO PREVENT ELECTRICAL SHOCK, CONNECT THE UNIT TO A PROPERLY GROUNDED
CAUTION
POWER OUTLET. DO NOT USE A THREE TO TWO PRONG ADAPTER. THIS DEFEATS SAFETY
GROUNDING.
DO NOT WEAR THE ESD WRIST STRAP WHEN YOU WORK ON LIVE CIRCUITS AND MORE
CAUTION
THAN 30 V PEAK IS PRESENT.
3 - 2
11
Average Installation Time
Average
Installation
Time
0.5 hour
4 hours
2 hours
Vivid 7 Acclimate Time
KEEP STORAGE TEMPERATURE WITHIN THESE LIMITS
-20
-15
-10
-5
0
-4
5
14
23
32
12
10
8
6
4
Section 3-2 - Installation Reminders
Comments
Dependent on the configuration
Dependent on the configuration
5
10
15
20
25
41
50
59
68
77
2
0
0
0
0
V
7 S
IVID
ERVICE
30
35
40
45
50
86
95
104
113
122
0
0
0
2
4
M
ANUAL
55
60
131
140
6
8

Advertisement

Table of Contents
loading