Mounting O 2 Compressed Gas Cylinders To; The Trolley - Dräger Savina 300 Instructions For Use Manual

Hide thumbs Also See for Savina 300:
Table of Contents

Advertisement

Assembly and preparation
Mounting O
compressed gas cylinders
2
to the trolley
Prerequisites:
– Gas cylinder holder option is available.
– Compressed gas cylinders have the following
dimensions:
Diameter
Trolley
80 to 176 mm
(3.15 to
6.93 in)
Compact
85 to 106 mm
trolley
(3.35 to
4.17 in)
WARNING
Risk of personal injury and damage to the
device
If the compressed gas cylinders are not
securely fastened to the trolley, they can fall
down.
Securely attach the compressed gas cylinders
to the trolley using both hook-and-loop straps.
WARNING
Risk of personal injury and damage to the
device
If the pressure reducers protrude beyond the
device, they may be damaged during trans-
port.
Position the compressed gas cylinders in a
way that prevents damage being caused to the
pressure reducers.
46
Trolley for Savina 300
Length incl.
pressure
reducer
420 to 760 mm
(16.54 to
29.92 in)
290 to 535 mm
(11.42 to
21.06 in)
A Pressure reducer
B Hose holder
C Hook-and-loop straps
Mount the compressed gas cylinders:
1 Place the compressed gas cylinders into the
2 Secure each compressed gas cylinder with
3 Secure the compressed gas hoses by hanging
A
C
C
mountings on the trolley.
2 hook-and-loop straps (C). If required, have
service personnel perform the following
adjustments:
 Adjust the height of the upper gas cylinder
holder to the compressed gas cylinders to
be used. The height must be adjusted so
that the top half of the compressed gas
cylinders is held firmly in place by the upper
cylinder holder.
 Replace the hook-and-loop straps. The
length of the hook-and-loop straps must
match the circumference of the compressed
gas cylinders.
them over the hose holders (B).
Instructions for use Savina 300 SW 5.n
A
B
C
B
C

Advertisement

Table of Contents
loading

Table of Contents