Checking Compact Breathing System; In Addition - Dräger Primus Series Quick Manual

Table of Contents

Advertisement

Preparation for Use
Checking Compact Breathing
System
Ensure that the compact
breathing system (1) is connected
and the ventilator is locked in place.
Check status of CO
absorbent (2), breathing bag
and patient hoses.

In Addition ...

Test the oxygen flush (3).
Check oxygen safety control (4).
Make sure vaporisers are in
position, locked and filled (5).
Check water trap level and the
sample gas line connection (6).
Check function of suction
system (7). Your suction system may
differ from the system shown.
Connect anaesthetic gas
scavenging system.
5
6
2
7
1 34